Description:
- The RM-044 is a low-cost, stand-alone instrument designed to offer signal conditioning for strain gauge type transducers such as pressure transducers and load cell.
- Supports Stain Gauge Work: ½ Bridge and ¼ Bridge 120- and 350-Ohm Strain Gauges. Internal Bridge completion provided.
- Gain and gross offset correction are both adjustable, as are all internal links.
- The electronics are enclosed in a 100 x 115 x 20 mm polycarbonate enclosure.
- Four 3-pin connectors are supplied for signal, transducer, and power connections.
- The RM-044 typically outputs +10 volts across the transducer's whole range, but it can also drive over a ±10-volt range.
- If necessary, the output can generate signals ranging from 4 to 20 mA.
Specifications:
- Power Supply: +24 Volt Nominal. (18 To 30 V). (Optional 12 V).
- Transducer Excitation: +1.3 -> +10 VOLT @ 50 mA MAX. (Or +/- 15 Volt unregulated @ 15 mA Max. Link selectable).
- Excitation Stability: 0.02 % / °C.
- Input Impedance: > 100 M ohm (Differential).
- CMRR: 100 dB @ 50 Hz.
- Linearity: 0.01 %.
- Gain Range: 120 TO 3200.
- Frequency Response: 10 Hz Or 20 kHz 4 Pole Max Flat (Factory Configured).
- Outputs: ± 10 VOLTS @ 5 mA MAX. 4 TO 20 mA (500-ohm MAX).
- Output noise: 1 mV RMS.